Artifact sources in ECG monitoring

Thank You for visiting this page in web-site section "24h ECG (Holter) monitoring" for students of medicine.

In ECG, stress testing, event monitoring and Holter recording, artifact is loosely defined as electrical interference or, “noise,” that is recorded from sources other than the electronic signals of the heart.  Artifact is a distortion of the signal being recorded.  Artifact or extraneous electronic signals can obscure the very information cardiology professionals are trying to obtain. The two main types of artifact are:

60 cycle (60 Hz) artifact in ECG monitoring

  • 60 cycle (60 Hz) noise that appears as, “fuzz,” all over the signal;   

Wandering baseline artifactt in ECG monitoring

  • Wandering baseline seen as an undulating, low frequency wave the ECG rides on.

There are a variety of sources for artifact in Holter recordings.  Although the 60 cycle noise can be caused by skeletal muscle contractions and a moderate amount of wandering baseline may be due to respiration, much of the artifact encountered in Holter recordings can be reduced or eliminated.  Female and pediatric patients must be treated the same as males to improve recording quality well enough to evaluate the subtle components of ECG such as P-wave, P-R interval, QRS duration and S-T segment for early detection of cardiac abnormalities.

Although the patient can cause artifact in the recording of an ECG, this may not be under the control of the cardiac technician performing the test.  Skeletal muscle contractions are a common source of brief 60 cycle artifacts in any ECG.  The electrical activity of skeletal muscle and that of the heart muscle are similar.  Muscle noise due to movement can be reduced, but not eliminated.  As this noise is generally of short duration, it is not our main concern.  If the subject being monitored is planning prolonged, strenuous activity or is performing a stress test, lead location and skin preparation guidelines below are even more important.  If the patient scratches at the electrodes, pulls on the lead wires or interferes with the recorder, artifacts will result and the test may be invalid.  As patient cooperation is very important, it is essential to communicate the needs required for a successful test with the patient.

Some sources of 60 cycle noise are common mechanical faults such as broken lead wires, poor quality electrodes or electrodes that are not compatible with the snap lead.  Because Event and Holter recorders do receive some physical abuse during use, loose or broken components need to be replaced.  The snap lead connector and/or the patient cable connection to the recorder should be stable.  The size of the electrode is not as important as long as they are Silver/Silver Chloride electrodes with plenty of adhesive to last 24-48 hours (or more for Event monitoring).  If the recording is going to be of a longer duration, the patient should be supplied with extra electrodes and instructed on the best way to remove the old and replace the new ones.  It is also important to check that the snap lead is compatible with the electrode so that it is not too loose and able to become disconnected.  Assuming that the recorder, lead wires and electrodes are all compatible and in excellent condition, the following procedures will reduce artifact.

How do I use 24h-monitoring.com?

You can learn more about the subjects within 24h-monitoring.com by going to the subject list and clicking on a subject of interest. You are welcome to use all of the information gathered to date. If you find that any of the information provided is in error, outdated or needs more information added, please use the contact us screen or the social networks to send your comments or new material. We will gladly give credit for any original information we include. We also welcome your questions and comments on the subjects covered, our project or areas you would like to see included. Welcome to 24h-monitoring.com!